Electrode array features
|
|
|
- Active length of the electrode array: 25 mm.
|
- Base diameter: 0,9 mm (advised size of the cochleostomy).
|
|
|
- Extra-cochlear section diameter: 1,1 mm.
|
- Electrode surface: 0,8 mm2 to 1,8 mm2
|
|
|
- Short version Digisonic® SP SI (Short Insertion) for residual hearing:
12 electrodes - 15 mm insertion.
|
 |
Stimulation features
|
- 20 channels. Current or voltage stimulation.
|
- Monopolar/common-ground stimulation through biphasic pulses.
|
- Capacitive coupling on each of the 20 electrodes: averaged residual current less than 100 nA.
|
- Selectable pulse amplitude.
|
- Pulse duration: from 1 to 120 µs (resolution = 0,5 µs).
|
- Stimulation frequency: 24 000 pps (pulses per second).
|
- Electrode impedance less than 2 k?.
|
Objectives measures
|
- Objective measures with Electrical Auditory Brainstem Responses and
Digistim® SP system. Windows 98 SE, ME, 2000 and XP compatible.
|
- Electrode impedance measurement by telemetry.
|
Contents of Digisonic®SP
|
- 1 Digisonic®SP implant - 1 Silicone dummy implant
2 Screws - 1 Owner card - 1 Instruction for use - 1 Implantation form
1 Explantation form - 1 Warranty certificate.
|